Anaesthesia System KAM-A101 is equipped with color LCD screen for clear display of trends, dosage rate and other setting parameters. Designed to suit adult as well as pediatric patients. System complies with closed, semi-closed, semi-open or assisted ventilation. Integrated sensors and valves ensure risk prevention from system failure and increases patient safety.
| Drive | Electronically controlled, bellow driven |
|---|---|
| Tidal Volume | 0 ml – 1500 ml |
| Configuration | |
| Gas Configuration | O<sub>2</sub>, N<sub>2</sub>O |
| Flow Meter | 1 l/min – 10 l/min |
| LCD Display | 7 inch LCD |
| Graphic Display | |
| Vaporizer | Single [Isoflurane / Enflurane / Sevoflurane / Halothane] |
| Fresh Gas Compensation | 25 l/min – 75 l/min |
| I:E Range | 4:1-1:8 |
| Pressure Range | |
| P trigger Range | |
| Monitoring | VT, MV, BPM, Paw |
| Modes of Ventilation | IPPV, SIMV, SIPPV, Manual |
| Respiration range | 1 bpm – 100 bpm |
| Concentration Range | 0% to 5% |
| Calibration | Automatic |
| Pipeline Input | |
| O2 Flush | |
| Minute Volume | |
| PEEP | |
| Run-time | |
| Waveforms | Pressure-time, Flow-time, Pressure-volume loop |
| Hypoxic Safety System | N<sub>2</sub>O cut-off valve / O<sub>2</sub> > 25% |
| Alarm | AC power failure, minute volume too high (low), airway pressure too high ( low ) and insufficiency of O<sub>2</sub> and N<sub>2</sub>O |
| Pressure Safety Valve | < 12.5 kPa |
| Volume of CO2 absorber | 1.5 L |
| Output Pressure | |
| Power Consumption | 50 VA |
| Power Supply | 100 VAC – 240 VAC, 50 / 60 Hz |
| Battery | [Rechargeable Lead-acid battery] </br> 30 minutes ~ 120 minutes |
| Current | |
| Working Condition | Patient airway pressure range: - 2 kPa ~ 10 kPa </br> Airway monitoring range: 0 ~ 9.9 kPa </br> Pressure limit range: 2.0 kPa ~ 6.0 kPa |
| Dimensions | 1190 x 410 x 710 mm |
| Weight | 82 Kgs |
Used in Surgical or Gynaecological departments to administer appropriate anaesthesia dosage at a specified rate to ensure effective sedation during surgeries and other invasive diagnostic or interventional procedures.